Membranes Part 2: Characteristics and Importance in Point-of-Care Diagnostic Assays
Hans Boehringer, Ph.D., and Bob Hudak, Co-Founders, Artemis Dx
,
In the first part of discussing the use of nitrocellulose membranes in lateral flow assays, we focused on the characteristics and
importance of the membranes. In part two, we elaborate on the selection of the best membrane type for a point-of-care lateral
flow test. The choice of membrane significantly influences assay performance, including sensitivity, specificity, reproducibility,
and shelf life. Selecting the appropriate nitrocellulose membrane requires a careful evaluation of multiple factors such as pore size,
flow rate, protein binding capacity, backing material, and compatibility with detection methods and particles.
1. Flow Rate and Pore Size
Flow rate, often inversely related to pore size, is a primary consideration in membrane selection. Membranes with slower flow
rates (e.g., 150-180 seconds/4 cm) allow for prolonged interaction time between analytes and capture antibodies, improving
sensitivity. Conversely, faster membranes (e.g., 60-95 seconds/4 cm) can reduce assay run time and are suitable when high analyte
concentrations are expected. Pore size, a measurement used in unbacked membranes typically ranging from 6 to 15 microns, affects
both flow dynamics and binding surface area. Smaller pores increase surface area and binding capacity but will prolong the flow rate
and increase background noise, while larger pores are allowing faster flow but lower the protein binding capacity,
2. Influence of Sample Type on Nitrocellulose Membrane Selection
The nature of the sample-whether it's whole blood, serum, respiratory or vaginal swabs, urine, saliva, or environmental
extract-greatly impacts nitrocellulose membrane selection in lateral flow assays. Each sample type has unique viscosity, particulate
content, and biomolecular composition, influencing flow dynamics and membrane interactions. Several manufacturers, including
Cytiva (Marlborough, MA), Sartorius Stedim (Bohemia, NY), and MilliporeSigma (Burlington, MA), provide technical information and
membrane selection guides when using different samples.
For viscous or particulate-rich samples (e.g., saliva, swabs, sputum, fecal samples, food extracts), membranes with larger pore
sizes and higher flow rates help prevent clogging and maintain capillary flow. In contrast, low-viscosity samples like urine or
water based environmental samples allow for smaller pores sizes and slower flow rates that will enhance analytical sensitivity.
Biological matrices can also introduce nonspecific binding and background noise. In such cases, membranes with optimized surface
chemistry or pre-treatment (e.g., blocking agents) can reduce specific matrix effects. For blood-based assays, membranes may also
require blood separation sample application pads or specialized treatments to prevent hemolysis and membrane staining.
Additionally, sample volume and required sensitivity influence membrane selection. Low-analyte concentration samples benefit from
slower membranes with high protein-binding capacity, allowing more time for analyte-antibody interaction.
Ultimately, the sample type dictates key membrane parameters-flow rate, pore size, and surface treatment or membrane blocking-making
early compatibility testing with real matrices essential for robust assay development.
3. Protein Binding Capacity
Nitrocellulose membranes are known for their ability to immobilize proteins through hydrophobic and electrostatic interactions.
High protein binding capacity (100-200 µg/cm²) is desirable for assays requiring dense capture zones. However, excessive binding
can lead to high background signals and reduced specificity. It's essential to match membrane binding capacity to the assay's needs,
optimizing the concentration of capture reagents during strip preparation.
4. Backing Material and Membrane Handling
Nitrocellulose membranes in lateral flow are typically cast onto plastic, mylar backings to enhance mechanical stability and ease
of manufacturing. The choice of backing material affects downstream processing, including cutting, lamination, and dispensing. A
membrane must also exhibit low variability in properties such as thickness, capillary flow rate, and surface uniformity to ensure
consistent assay performance across production batches.
5. Compatibility with Detection Systems
The detection method (e.g., colloidal gold, latex beads, fluorescent labels) also influences membrane choice. For instance,
fluorescent labels often require that membranes have low background fluorescence and minimal autofluorescence. When using
colorimetric labels, optical clarity and contrast of the membrane surface are important for visual or instrument-based
detection.
6. Environmental Stability and Storage
The membrane should be stable under the assay's storage and operational conditions. Some nitrocellulose membranes are sensitive
to humidity and temperature fluctuations, which can affect protein stability and membrane performance. Proper storage (e.g.,
desiccation, temperature control) and pretreatment (e.g., blocking, surfactant treatment) can mitigate these issues, but membrane
selection should consider environmental robustness for point-of-care applications.
7. Lot-to-Lot Consistency and Supplier Support
Reliable assay development requires membrane consistency. Variability between lots can undermine assay reproducibility and
regulatory compliance. Working with reputable suppliers who provide tight lot specifications, detailed quality control data, and
technical support is crucial for long-term success.
Conclusion
Selecting the right nitrocellulose membrane for a lateral flow assay involves a balance between
performance, manufacturability, and detection requirements. By carefully evaluating flow characteristics, binding capacity,
compatibility, and consistency, developers can optimize assay sensitivity, reliability, and ease of production-ultimately
ensuring a successful diagnostic product.